Study Investigates the Impact of Automating Respiration Rate Measurement Using Masimo Rad-G™ with RRp®

Masimo (NASDAQ: MASI) today announced the results of a prospective, observational study published in Acta Paediatrica in which researchers from the Hospital for Sick Children in Toronto evaluated the accuracy of plethysmographic respiration rate measurement (RRp® ) using Masimo Rad-G™, a rugged, handheld device, on malnourished, hospitalized children in Nigeria.1

This press release features multimedia. View the full release here: https://www.businesswire.com/news/home/20210328005012/en/

Noting that in resource-limited environments, respiration rate (RR) measurement is often used to directly inform medical decisions for children with respiratory problems, but that manual RR counting “remains a challenge,” Dr. Nancy Dale and colleagues investigated whether a technological solution might provide a useful alternative to manual counting. To make the evaluation, the researchers compared simultaneous device measurements and nurse-measured manual RR counts on malnourished children. The device chosen was the Masimo Rad-G, which uses a pulse oximetry sensor to measure both oxygen saturation and RRp, and which has been shown to provide good agreement between RRp and pediatrician-measured RR.2 They enrolled 514 children, aged 6 to 59 months, who were hospitalized between July 2019 and May 2020, in Borno State, Nigeria. Study nurses were trained to operate Rad-G and also perform manual RR counts as part of twice-daily patient assessment. RR was manually counted for 60 seconds while Rad-G simultaneously measured RRp via a sensor attached to the patient’s toe, and both measurements were recorded.

Analyzing the 6,889 paired RR measurements, the researchers found that the mean Rad-G RRp reading was 1.3 bpm (95% confidence interval 1.2 – 1.4 bpm) higher than the mean manual RR value. The mean absolute difference between the two methods was 4.4 bpm (95% CI 4.3 – 4.5 bpm). When RR was classified as either “normal” or “fast” breathing (using WHO pneumonia thresholds), the two methods resulted in the same classification 84% of the time. When RR was classified according to BedsidePEWS RR sub-scoring (a 4-point scale), 80% of the scores were the same, and 99.3% were within 1 point.

The researchers concluded that their findings “highlight the potential clinical impact of changing practice from manual to automated RR count. Clinical implementation of the device should be carefully monitored to measure impact on patient outcomes.”

Study co-author Dr. Stanley Zlotkin commented, “Technical solutions to improve clinical care are laudable. We look forward to continuing this research.”

RRp is one of multiple RR monitoring modalities offered by Masimo, which also include acoustic respiration rate (RRa® ) and NomoLine® capnography (RRc™), helping clinicians ensure they have the most suitable tool for each patient scenario.

First developed in partnership with The Bill & Melinda Gates Foundation, Rad-G is a rugged, handheld device that provides clinically proven Masimo SET® pulse oximetry, respiration rate (RRp), and other important parameters. With its long-lasting rechargeable battery, robust rubber casing, and light weight, Rad-G makes it easier for clinicians to quickly assess patients and make informed care decisions anywhere pulse oximetry or vital signs checking is needed in a compact, portable form factor. Coupled with the universal Mini-Clip™ pulse oximeter sensor to provide the ultimate in handheld versatility, Rad-G can be used in a variety of settings, including limited-resource environments, both indoors and in the field.

In the U.S., RRp is 510(k) cleared for patients greater than two years old.

@Masimo | #Masimo

About Masimo

Masimo (NASDAQ: MASI) is a global medical technology company that develops and produces a wide array of industry-leading monitoring technologies, including innovative measurements, sensors, patient monitors, and automation and connectivity solutions. Our mission is to improve patient outcomes and reduce the cost of care. Masimo SET® Measure-through Motion and Low Perfusion™ pulse oximetry, introduced in 1995, has been shown in over 100 independent and objective studies to outperform other pulse oximetry technologies.3 Masimo SET® has also been shown to help clinicians reduce severe retinopathy of prematurity in neonates,4 improve CCHD screening in newborns,5 and, when used for continuous monitoring with Masimo Patient SafetyNet™ in post-surgical wards, reduce rapid response team activations, ICU transfers, and costs.6-9 Masimo SET® is estimated to be used on more than 200 million patients in leading hospitals and other healthcare settings around the world,10 and is the primary pulse oximetry at 9 of the top 10 hospitals according to the 2020-21 U.S. News and World Report Best Hospitals Honor Roll.11 Masimo continues to refine SET® and in 2018, announced that SpO2 accuracy on RD SET® sensors during conditions of motion has been significantly improved, providing clinicians with even greater confidence that the SpO2 values they rely on accurately reflect a patient’s physiological status. In 2005, Masimo introduced rainbow® Pulse CO-Oximetry technology, allowing noninvasive and continuous monitoring of blood constituents that previously could only be measured invasively, including total hemoglobin (SpHb® ), oxygen content (SpOC™), carboxyhemoglobin (SpCO® ), methemoglobin (SpMet® ), Pleth Variability Index (PVi® ), RPVi™ (rainbow® PVi), and Oxygen Reserve Index (ORi™). In 2013, Masimo introduced the Root® Patient Monitoring and Connectivity Platform, built from the ground up to be as flexible and expandable as possible to facilitate the addition of other Masimo and third-party monitoring technologies; key Masimo additions include Next Generation SedLine® Brain Function Monitoring, O3® Regional Oximetry, and ISA™ Capnography with NomoLine® sampling lines. Masimo’s family of continuous and spot-check monitoring Pulse CO-Oximeters® includes devices designed for use in a variety of clinical and non-clinical scenarios, including tetherless, wearable technology, such as Radius-7® and Radius PPG™, portable devices like Rad-67™, fingertip pulse oximeters like MightySat® Rx, and devices available for use both in the hospital and at home, such as Rad-97® . Masimo hospital automation and connectivity solutions are centered around the Masimo Hospital Automation™ platform, and include Iris® Gateway, iSirona™, Patient SafetyNet, Replica™, Halo ION™, UniView™, UniView :60™, and Masimo SafetyNet™. Additional information about Masimo and its products may be found at www.masimo.com . Published clinical studies on Masimo products can be found at www.masimo.com/evidence/featured-studies/feature/ .

ORi and RPVi have not received FDA 510(k) clearance and are not available for sale in the United States. The use of the trademark Patient SafetyNet is under license from University HealthSystem Consortium.

References

  1. Dale N, Parshuram C, Tomlinson G, Shepherd S, Ashir GM, Maryah LB, Zlotkin S. Performance of automated versus nurse-measured respiratory rate measurements in hospitalized malnourished children. Acta Paediatr 2021. DOI: https://doi.org/10.1111/apa.15781 .
  2. Alwadhi V, Sarin E, Kuma P et al. Measuring accuracy of plethysmography based respiratory rate measurement using pulse oximeter at a tertiary hospital in India. Pneumonia 2020;12:4. DOI: https://doi.org/10.1186/s41479-020-00067-2 .
  3. Published clinical studies on pulse oximetry and the benefits of Masimo SET® can be found on our website at http://www.masimo.com . Comparative studies include independent and objective studies which are comprised of abstracts presented at scientific meetings and peer-reviewed journal articles.
  4. Castillo A et al. Prevention of Retinopathy of Prematurity in Preterm Infants through Changes in Clinical Practice and SpO2 Technology. Acta Paediatr. 2011 Feb;100(2):188-92.
  5. de-Wahl Granelli A et al. Impact of pulse oximetry screening on the detection of duct dependent congenital heart disease: a Swedish prospective screening study in 39,821 newborns. BMJ. 2009;Jan 8;338.
  6. Taenzer A et al. Impact of pulse oximetry surveillance on rescue events and intensive care unit transfers: a before-and-after concurrence study. Anesthesiology . 2010:112(2):282-287.
  7. Taenzer A et al. Postoperative Monitoring – The Dartmouth Experience. Anesthesia Patient Safety Foundation Newsletter . Spring-Summer 2012.
  8. McGrath S et al. Surveillance Monitoring Management for General Care Units: Strategy, Design, and Implementation. The Joint Commission Journal on Quality and Patient Safety . 2016 Jul;42(7):293-302.
  9. McGrath S et al. Inpatient Respiratory Arrest Associated With Sedative and Analgesic Medications: Impact of Continuous Monitoring on Patient Mortality and Severe Morbidity. J Patient Saf. 2020 14 Mar. DOI: 10.1097/PTS.0000000000000696.
  10. Estimate: Masimo data on file.
  11. http://health.usnews.com/health-care/best-hospitals/articles/best-hospitals-honor-roll-and-overview .

Forward-Looking Statements

This press release includes forward-looking statements as defined in Section 27A of the Securities Act of 1933 and Section 21E of the Securities Exchange Act of 1934, in connection with the Private Securities Litigation Reform Act of 1995. These forward-looking statements include, among others, statements regarding the potential effectiveness of RRp® , Rad-G™, and SET® . These forward-looking statements are based on current expectations about future events affecting us and are subject to risks and uncertainties, all of which are difficult to predict and many of which are beyond our control and could cause our actual results to differ materially and adversely from those expressed in our forward-looking statements as a result of various risk factors, including, but not limited to: risks related to our assumptions regarding the repeatability of clinical results; risks related to our belief that Masimo's unique noninvasive measurement technologies, including RRp, Rad-G, and SET® , contribute to positive clinical outcomes and patient safety; risks related to our belief that Masimo noninvasive medical breakthroughs provide cost-effective solutions and unique advantages; risks related to COVID-19; as well as other factors discussed in the "Risk Factors" section of our most recent reports filed with the Securities and Exchange Commission ("SEC"), which may be obtained for free at the SEC's website at www.sec.gov . Although we believe that the expectations reflected in our forward-looking statements are reasonable, we do not know whether our expectations will prove correct. All forward-looking statements included in this press release are expressly qualified in their entirety by the foregoing cautionary statements. You are cautioned not to place undue reliance on these forward-looking statements, which speak only as of today's date. We do not undertake any obligation to update, amend or clarify these statements or the "Risk Factors" contained in our most recent reports filed with the SEC, whether as a result of new information, future events or otherwise, except as may be required under the applicable securities laws.

Link:

ClickThru

About Business Wire

Business Wire
Business Wire
101 California Street, 20th Floor
CA 94111 San Francisco

http://businesswire.com

Subscribe to releases from Business Wire

Subscribe to all the latest releases from Business Wire by registering your e-mail address below. You can unsubscribe at any time.

Latest releases from Business Wire

Multi-Color Corporation Initiates Implementation of Restructuring Agreement29.1.2026 07:58:00 CET | Press release

Commences Prepackaged Chapter 11 Process As Part of Previously Announced Restructuring Support Agreement to Reset Balance Sheet and Position Company for Long-Term Growth and InvestmentRestructuring Supported by CD&R and More Than Supermajority of Senior Secured Lenders Who Have Agreed to Backstop Nearly $890 Million Investment; Provides More Than $500 Million of New Liquidity Upon EmergenceAll Global Operations and Service to Customers Expected to Continue Without Interruption; All Trade Vendors and Suppliers Expected to be Paid in FullProvides for $250 Million of New Money Debtor-in-Possession Financing to Capitalize the Business Throughout the Prepackaged Chapter 11 Process Multi-Color Corporation (“MCC” or the “Company”), a global leader in prime label solutions, today announced that to implement the previously announced restructuring support agreement (the “RSA”), the Company has commenced its prepackaged Chapter 11 filing in the United States Bankruptcy Court for the District of N

Takeda Reports Third-Quarter FY2025 Results: Updates Full Year Outlook to Reflect VYVANSE® Generics Impact, OPEX Discipline and FX Tailwind; Progressing Toward Three Transformative Launches Ahead29.1.2026 07:37:00 CET | Press release

Year-to-Date Revenue Declined by 2.8% at Constant Exchange Rate (CER), 3.3% atActual Exchange Rates (AER);Impact of VYVANSE® Generics is Tapering OffCore Operating Profit Declined by 3.4% at both CER and AER Year to DateReported Operating Profit Increased by 1.2% at AER Year to Date, Lower Restructuring Expenses More Than Offset Impairment Takeda (TOKYO:4502/NYSE:TAK) today announced earnings results for the third quarter of fiscal year 2025 (nine months ended December 31, 2025). The gap between incremental Growth & Launch Products revenue and VYVANSE erosion is narrowing, and operational efficiencies drove year-on-year reductions in operating expenses, including R&D. The company raised its full-year forecasts based on cost discipline and FX tailwind, while its Revenue Management Guidance has been updated primarily due to the impact of VYVANSE generics. Takeda is positioned for long-term growth and has multiple late-stage programs with multibillion-dollar peak revenue potential. Follow

Galderma Unveils ‘Wake Up To Restylane’, Underscoring Restylane as the Ideal Hyaluronic Acid Treatment for Always-On Natural Beauty29.1.2026 07:00:00 CET | Press release

‘Wake Up To Restylane®’ is a global campaign designed to meet the rising demand for natural-looking always-on results, helping people wake up looking refreshed and rested, and reframes Restylane treatments from a clinical procedure into an always-on beauty regimen Backed by 30 years of science, the Restylane portfolio is the only hyaluronic acid (HA) range offering four distinct technologies, designed to deliver personalized natural-looking outcomes that provide contour, definition, and hydration, with HA closest to the skin’s own1-7 This launch builds on new clinical data presented at the International Master Course on Aging Science (IMCAS) World Congress 2026, confirming Restylane’s proven efficacy and versatility across multiple facial areas and the décolletage8-11 Galderma (SIX: GALD), the pure-play dermatology category leader, has launched ‘Wake Up To Restylane’, a global campaign that showcases Restylane as an everyday beauty ally. Aligned with the growing desire for effortless,

Ubitus Partners with Maizuru City, Kyoto to Launch AI Data Center Project, Advancing the Deployment of a Top-Tier AI GPU Center in Japan29.1.2026 07:00:00 CET | Press release

Ubitus K.K. (Headquarters: Shinjuku, Tokyo; Representative Director & CEO: Wesley Kuo), a global leader in cloud streaming and AI solutions, today announced that it will hold a land signing ceremony and press conference with Maizuru City, Kyoto Prefecture, on January 29, 2026. At the event, Ubitus will officially unveil its comprehensive plan to construct an AI Data Center in Maizuru City, marking the company’s transition into the physical construction phase of its initiative to build a top-tier AI GPU center in Japan. This press release features multimedia. View the full release here: https://www.businesswire.com/news/home/20260128944842/en/ Ubitus will launch its Maizuru City AI Data Center project on January 29, marking the first step toward building a top-tier AI GPU center in Japan. Powered by NVIDIA Blackwell GPUs and NeoCloud, the project strengthens Japan’s AI infrastructure and regional development. This land agreement represents the first major milestone following Ubitus’ sel

Lattice to Showcase Advanced Edge AI Solutions at the FPGA-forum 202628.1.2026 22:00:00 CET | Press release

Lattice Semiconductor (NASDAQ: LSCC), the low power programmable leader, today announced its exhibition plan for the upcoming FPGA-forum 2026 taking place February 11 – 12, 2026 in Trondheim, Norway. As part of the event, Lattice Corporate Vice President of Software Solutions and Applications Engineering Eleena Ong will deliver a keynote presentation exploring how rapidly maturing AI technologies can turn FPGA platforms into innovation canvases for a much broader range of developers, sparking new application possibilities. Lattice will also deliver a technical presentation and host a demo showcase focused on how its low power FPGA solutions are advancing connectivity and edge AI applications. Who: Lattice Semiconductor What / When (GMT+1): Lattice Demo Showcase (Table #3), Feb 11-12 Keynote Feb. 11, 9:30 a.m. “Unlocking the Next Wave of FPGA Innovators with Generative and Agentic AI” Technical Presentation Feb. 12, 10 a.m. “Smaller. Cooler. Smarter: Lattice FPGAs’ Path to Uncompromised

In our pressroom you can read all our latest releases, find our press contacts, images, documents and other relevant information about us.

Visit our pressroom
World GlobeA line styled icon from Orion Icon Library.HiddenA line styled icon from Orion Icon Library.Eye