Study Investigates the Impact of Automating Respiration Rate Measurement Using Masimo Rad-G™ with RRp®

Masimo (NASDAQ: MASI) today announced the results of a prospective, observational study published in Acta Paediatrica in which researchers from the Hospital for Sick Children in Toronto evaluated the accuracy of plethysmographic respiration rate measurement (RRp® ) using Masimo Rad-G™, a rugged, handheld device, on malnourished, hospitalized children in Nigeria.1

This press release features multimedia. View the full release here: https://www.businesswire.com/news/home/20210328005012/en/

Noting that in resource-limited environments, respiration rate (RR) measurement is often used to directly inform medical decisions for children with respiratory problems, but that manual RR counting “remains a challenge,” Dr. Nancy Dale and colleagues investigated whether a technological solution might provide a useful alternative to manual counting. To make the evaluation, the researchers compared simultaneous device measurements and nurse-measured manual RR counts on malnourished children. The device chosen was the Masimo Rad-G, which uses a pulse oximetry sensor to measure both oxygen saturation and RRp, and which has been shown to provide good agreement between RRp and pediatrician-measured RR.2 They enrolled 514 children, aged 6 to 59 months, who were hospitalized between July 2019 and May 2020, in Borno State, Nigeria. Study nurses were trained to operate Rad-G and also perform manual RR counts as part of twice-daily patient assessment. RR was manually counted for 60 seconds while Rad-G simultaneously measured RRp via a sensor attached to the patient’s toe, and both measurements were recorded.

Analyzing the 6,889 paired RR measurements, the researchers found that the mean Rad-G RRp reading was 1.3 bpm (95% confidence interval 1.2 – 1.4 bpm) higher than the mean manual RR value. The mean absolute difference between the two methods was 4.4 bpm (95% CI 4.3 – 4.5 bpm). When RR was classified as either “normal” or “fast” breathing (using WHO pneumonia thresholds), the two methods resulted in the same classification 84% of the time. When RR was classified according to BedsidePEWS RR sub-scoring (a 4-point scale), 80% of the scores were the same, and 99.3% were within 1 point.

The researchers concluded that their findings “highlight the potential clinical impact of changing practice from manual to automated RR count. Clinical implementation of the device should be carefully monitored to measure impact on patient outcomes.”

Study co-author Dr. Stanley Zlotkin commented, “Technical solutions to improve clinical care are laudable. We look forward to continuing this research.”

RRp is one of multiple RR monitoring modalities offered by Masimo, which also include acoustic respiration rate (RRa® ) and NomoLine® capnography (RRc™), helping clinicians ensure they have the most suitable tool for each patient scenario.

First developed in partnership with The Bill & Melinda Gates Foundation, Rad-G is a rugged, handheld device that provides clinically proven Masimo SET® pulse oximetry, respiration rate (RRp), and other important parameters. With its long-lasting rechargeable battery, robust rubber casing, and light weight, Rad-G makes it easier for clinicians to quickly assess patients and make informed care decisions anywhere pulse oximetry or vital signs checking is needed in a compact, portable form factor. Coupled with the universal Mini-Clip™ pulse oximeter sensor to provide the ultimate in handheld versatility, Rad-G can be used in a variety of settings, including limited-resource environments, both indoors and in the field.

In the U.S., RRp is 510(k) cleared for patients greater than two years old.

@Masimo | #Masimo

About Masimo

Masimo (NASDAQ: MASI) is a global medical technology company that develops and produces a wide array of industry-leading monitoring technologies, including innovative measurements, sensors, patient monitors, and automation and connectivity solutions. Our mission is to improve patient outcomes and reduce the cost of care. Masimo SET® Measure-through Motion and Low Perfusion™ pulse oximetry, introduced in 1995, has been shown in over 100 independent and objective studies to outperform other pulse oximetry technologies.3 Masimo SET® has also been shown to help clinicians reduce severe retinopathy of prematurity in neonates,4 improve CCHD screening in newborns,5 and, when used for continuous monitoring with Masimo Patient SafetyNet™ in post-surgical wards, reduce rapid response team activations, ICU transfers, and costs.6-9 Masimo SET® is estimated to be used on more than 200 million patients in leading hospitals and other healthcare settings around the world,10 and is the primary pulse oximetry at 9 of the top 10 hospitals according to the 2020-21 U.S. News and World Report Best Hospitals Honor Roll.11 Masimo continues to refine SET® and in 2018, announced that SpO2 accuracy on RD SET® sensors during conditions of motion has been significantly improved, providing clinicians with even greater confidence that the SpO2 values they rely on accurately reflect a patient’s physiological status. In 2005, Masimo introduced rainbow® Pulse CO-Oximetry technology, allowing noninvasive and continuous monitoring of blood constituents that previously could only be measured invasively, including total hemoglobin (SpHb® ), oxygen content (SpOC™), carboxyhemoglobin (SpCO® ), methemoglobin (SpMet® ), Pleth Variability Index (PVi® ), RPVi™ (rainbow® PVi), and Oxygen Reserve Index (ORi™). In 2013, Masimo introduced the Root® Patient Monitoring and Connectivity Platform, built from the ground up to be as flexible and expandable as possible to facilitate the addition of other Masimo and third-party monitoring technologies; key Masimo additions include Next Generation SedLine® Brain Function Monitoring, O3® Regional Oximetry, and ISA™ Capnography with NomoLine® sampling lines. Masimo’s family of continuous and spot-check monitoring Pulse CO-Oximeters® includes devices designed for use in a variety of clinical and non-clinical scenarios, including tetherless, wearable technology, such as Radius-7® and Radius PPG™, portable devices like Rad-67™, fingertip pulse oximeters like MightySat® Rx, and devices available for use both in the hospital and at home, such as Rad-97® . Masimo hospital automation and connectivity solutions are centered around the Masimo Hospital Automation™ platform, and include Iris® Gateway, iSirona™, Patient SafetyNet, Replica™, Halo ION™, UniView™, UniView :60™, and Masimo SafetyNet™. Additional information about Masimo and its products may be found at www.masimo.com . Published clinical studies on Masimo products can be found at www.masimo.com/evidence/featured-studies/feature/ .

ORi and RPVi have not received FDA 510(k) clearance and are not available for sale in the United States. The use of the trademark Patient SafetyNet is under license from University HealthSystem Consortium.

References

  1. Dale N, Parshuram C, Tomlinson G, Shepherd S, Ashir GM, Maryah LB, Zlotkin S. Performance of automated versus nurse-measured respiratory rate measurements in hospitalized malnourished children. Acta Paediatr 2021. DOI: https://doi.org/10.1111/apa.15781 .
  2. Alwadhi V, Sarin E, Kuma P et al. Measuring accuracy of plethysmography based respiratory rate measurement using pulse oximeter at a tertiary hospital in India. Pneumonia 2020;12:4. DOI: https://doi.org/10.1186/s41479-020-00067-2 .
  3. Published clinical studies on pulse oximetry and the benefits of Masimo SET® can be found on our website at http://www.masimo.com . Comparative studies include independent and objective studies which are comprised of abstracts presented at scientific meetings and peer-reviewed journal articles.
  4. Castillo A et al. Prevention of Retinopathy of Prematurity in Preterm Infants through Changes in Clinical Practice and SpO2 Technology. Acta Paediatr. 2011 Feb;100(2):188-92.
  5. de-Wahl Granelli A et al. Impact of pulse oximetry screening on the detection of duct dependent congenital heart disease: a Swedish prospective screening study in 39,821 newborns. BMJ. 2009;Jan 8;338.
  6. Taenzer A et al. Impact of pulse oximetry surveillance on rescue events and intensive care unit transfers: a before-and-after concurrence study. Anesthesiology . 2010:112(2):282-287.
  7. Taenzer A et al. Postoperative Monitoring – The Dartmouth Experience. Anesthesia Patient Safety Foundation Newsletter . Spring-Summer 2012.
  8. McGrath S et al. Surveillance Monitoring Management for General Care Units: Strategy, Design, and Implementation. The Joint Commission Journal on Quality and Patient Safety . 2016 Jul;42(7):293-302.
  9. McGrath S et al. Inpatient Respiratory Arrest Associated With Sedative and Analgesic Medications: Impact of Continuous Monitoring on Patient Mortality and Severe Morbidity. J Patient Saf. 2020 14 Mar. DOI: 10.1097/PTS.0000000000000696.
  10. Estimate: Masimo data on file.
  11. http://health.usnews.com/health-care/best-hospitals/articles/best-hospitals-honor-roll-and-overview .

Forward-Looking Statements

This press release includes forward-looking statements as defined in Section 27A of the Securities Act of 1933 and Section 21E of the Securities Exchange Act of 1934, in connection with the Private Securities Litigation Reform Act of 1995. These forward-looking statements include, among others, statements regarding the potential effectiveness of RRp® , Rad-G™, and SET® . These forward-looking statements are based on current expectations about future events affecting us and are subject to risks and uncertainties, all of which are difficult to predict and many of which are beyond our control and could cause our actual results to differ materially and adversely from those expressed in our forward-looking statements as a result of various risk factors, including, but not limited to: risks related to our assumptions regarding the repeatability of clinical results; risks related to our belief that Masimo's unique noninvasive measurement technologies, including RRp, Rad-G, and SET® , contribute to positive clinical outcomes and patient safety; risks related to our belief that Masimo noninvasive medical breakthroughs provide cost-effective solutions and unique advantages; risks related to COVID-19; as well as other factors discussed in the "Risk Factors" section of our most recent reports filed with the Securities and Exchange Commission ("SEC"), which may be obtained for free at the SEC's website at www.sec.gov . Although we believe that the expectations reflected in our forward-looking statements are reasonable, we do not know whether our expectations will prove correct. All forward-looking statements included in this press release are expressly qualified in their entirety by the foregoing cautionary statements. You are cautioned not to place undue reliance on these forward-looking statements, which speak only as of today's date. We do not undertake any obligation to update, amend or clarify these statements or the "Risk Factors" contained in our most recent reports filed with the SEC, whether as a result of new information, future events or otherwise, except as may be required under the applicable securities laws.

Link:

ClickThru

About Business Wire

Business Wire
Business Wire
101 California Street, 20th Floor
CA 94111 San Francisco

http://businesswire.com

Subscribe to releases from Business Wire

Subscribe to all the latest releases from Business Wire by registering your e-mail address below. You can unsubscribe at any time.

Latest releases from Business Wire

IonQ and QuantumBasel Expand Long-Term Partnership in Next-Generation Quantum Systems17.12.2025 22:10:00 CET | Press release

Extension solidifies QuantumBasel as IonQ’s Innovation Center in Europe; adds IonQ Tempo and next-generation system to advance quantum commercialization IonQ (NYSE: IONQ), the world’s leading quantum company, today announced an expanded agreement with QuantumBasel, the quantum initiative of uptownBasel, Switzerland’s innovation campus. The extended contract grants QuantumBasel ownership of its existing IonQ Forte Enterprise system and secures ownership of a next-generation Tempo system. This new agreement brings the total deal value of the QuantumBasel and IonQ partnership to over $60 million and extends IonQ’s on-site presence in Switzerland four more years, continuing through 2029. QuantumBasel is IonQ’s official Innovation Center in Europe, serving as a hub for European industry, academia, and research institutions to explore practical quantum computing applications and access IonQ’s latest enterprise-grade systems. “Our extended partnership with QuantumBasel represents a cornerston

Suzano Starts Up New Production Line, Boosting Its Fluff Pulp Capacity by 400%17.12.2025 21:50:00 CET | Press release

A R$490 million investment expands the supply of raw material used in the production of absorbent items Suzano, the world’s largest pulp producer, has commenced operations this week at its new fluff pulp production line located in its Limeira unit in Brazil’s São Paulo state. This R$490 million investment increases Suzano’s total fluff pulp production capacity by more than 400%, from 100,000 to 440,000 tonnes per year. The project involved converting the existing pulp line at the Limeira unit into a flexible machine, capable of producing both Eucafluff® and market pulp. Eucafluff® is used in the production of absorbent and personal hygiene products, such as baby and adult diapers, sanitary pads and pet pads. Then market pulp is supplied for making products including toilet paper, printing and writing papers, and paper packaging. Launched in 2015, Eucafluff® is the world’s first fluff pulp made from eucalyptus, delivering unique advantages like enhanced softness and flexibility, which t

SES Acknowledges Moody’s Rating Action and Reiterates Deleveraging Commitments17.12.2025 21:36:00 CET | Press release

SES S.A. (“SES” or the “Company”), a leading space solutions company, acknowledges the credit rating action announced by Moody’s Investor Service today, which follows the release of SES’ Q3 2025 results and Intelsat integration update. SES management reiterates that the Company continues to execute on its strategy with a clear plan to strengthen its key credit metrics over time. Consistent with this plan, it remains management’s intention to de-lever and return to credit metrics that are commensurate with investment grade, with a policy objective of reducing adjusted net leverage1 to at least 3.0x or below. Today’s rating action does not change the Company’s ability to operate its business, serve customers, or execute its strategic plan. SES maintains a balanced weighted average debt maturity profile of approximately five years, and the rating action from Moody’s is not expected to have a material impact on the interest payable under the Company’s existing debt facilities. SES also ben

Picsart and Zazzle Power the Creator-to-Commerce Evolution with Print-On-Demand Integration17.12.2025 20:47:00 CET | Press release

Partnership Gives Creators an End-to-End Solution to Create A Range of Products in the $205 Billion Creator Economy Picsart, the world’s leading digital creation platform with over two billion downloads and more than 100 million monthly active users, is partnering with Zazzle, the global leader in customized products and designs, to deliver an end-to-end printing solution for Picsart users. The new “vibe-design-to-print” integration works seamlessly with Picsart’s recently expanded suite of AI products, including AI Assistant and Flows, enabling creators to ideate, iterate, design, and bring their work to life as physical art and products. This press release features multimedia. View the full release here: https://www.businesswire.com/news/home/20251217882832/en/ Picsart and Zazzle Power the Creator-to-Commerce Evolution with Print-On-Demand Integration “At Picsart, our mission has always been to give creators the tools they need to bring their ideas to life,” said Hovhannes Avoyan, Fo

Inspiring Leadership Recognised: Anna Byrne Named Finalist at CIPD 2026 HR Awards17.12.2025 19:58:00 CET | Press release

ProAmpac, a global leader in flexible packaging and material science, is proud to announce that Anna Byrne has been selected as a finalist in the “Most Inspirational and Impactful Leader” category at the CIPD 2026 HR Awards in Ireland. This recognition highlights Anna’s exceptional leadership, her unwavering commitment to people-first practices, and her meaningful influence across the organisation. This press release features multimedia. View the full release here: https://www.businesswire.com/news/home/20251217690526/en/ The CIPD HR Awards in Ireland The CIPD HR Awards highlight the people profession’s crucial role in driving strategy, leading transformation, and enabling organisations to thrive. Overseen by CIPD Ireland and supported by independent academic expertise, the awards celebrate innovation, integrity, and impact within modern HR practice. This year, the awards introduce four new categories, including the “Most Inspirational and Impactful Leader” award, reflecting the evolvi

In our pressroom you can read all our latest releases, find our press contacts, images, documents and other relevant information about us.

Visit our pressroom
World GlobeA line styled icon from Orion Icon Library.HiddenA line styled icon from Orion Icon Library.Eye