MASIMO
Masimo (NASDAQ: MASI) today announced the results of a prospective study published in the Sri Lanka Journal of Child Health in which researchers in Colombo, Sri Lanka evaluated the efficacy of a pulse oximetry-based critical congenital heart disease (CCHD) newborn screening strategy using Masimo SET® pulse oximetry.1 The authors concluded that pulse oximetry is a “simple, noninvasive, cost-effective, feasible, and reliable test,” and found that it had higher CCHD screening sensitivity than physical exam. Combining the two methods led to detection of all cases of CCHD in the study cohort, and they recommended that, “Pulse oximetry screening as a combined strategy with newborn physical exam should be implemented as a basic routine at discharge for every newborn in maternity units island-wide.” As they note, their work is the first published CCHD study of this nature in Sri Lanka.
This press release features multimedia. View the full release here: https://www.businesswire.com/news/home/20210322005184/en/
Noting that while “developed countries have abundant research on pulse oximetry screening” for CCHD, “there are few studies in developing countries,” Dr. CR Gunaratne and colleagues sought to study the utility of such a screening strategy in their local setting. From November 2018 to April 2019, researchers assessed the rate of detection of CCHD using Masimo SET® pulse oximetry compared to routine physical exam alone in 5,435 asymptomatic newborns admitted to the post-natal wards at Castle Street Hospital, Colombo. Physical exam was performed at ≥ 24 hours of age “to identify any visible central cyanosis, weak/absent femoral pulses or cardiac murmur” by an experienced medical officer, blinded to pulse oximetry results. Radical-7® Pulse CO-Oximeters® with Masimo SET® pulse oximetry were used to measure pre-ductal and post-ductal oxygen saturation (SpO2 ) on the right hand and right foot, respectively, as part of a standardized screening algorithm. For newborns with positive results, an echocardiogram was performed within 48 hours to diagnose CCHD.
The researchers found that Masimo SET® pulse oximetry had a CCHD detection rate of 91%, compared to 82% for physical exam. The addition of Masimo SET® pulse oximetry to physical exam screening led to the detection of 2 cases missed by physical exam alone, with a combined detection rate of 100%. The positive predictive value and positive likelihood ratio were both higher for SET® pulse oximetry compared to physical exam (71.4% vs. 8.6% and 1232.7 vs. 46.2, p = 0.0001). The researchers also found that the false positive rate was “substantially” lower for SET® pulse oximetry compared to physical exam (0.07% vs. 1.76%, p = 0.0001).
The researchers concluded, “Prevalence of CCHD in our study was 2.02 per 1000 live births. Using a pulse oximetry strategy as an adjunct to routine physical exam can substantially reduce the diagnostic gap in CCHD as [a] combined approach has an additive effect resulting in more efficient screening.”
Since its introduction in 1995, Masimo Measure-through Motion and Low Perfusion™ Signal Extraction Technology® (SET® ) has been shown in more than 100 independent and objective studies to outperform other pulse oximetry technologies, providing clinicians with increased sensitivity and specificity to help them make critical patient care decisions.2 To date, nine other published CCHD screening studies, all with positive conclusions and representing over 300,000 infants, have used Masimo SET® ,3-11 which includes the largest CCHD study to date, of 122,738 newborns.5 All of the CCHD studies with Masimo SET® pulse oximetry have shown improved screening sensitivity with the use of Masimo SET® alongside clinical assessment when compared to routine physical exam alone. Results from CCHD studies using other pulse oximetry technologies have shown that other technologies do not offer the same performance as Masimo SET® during CCHD screening.12-14
With its ability to accurately measure through motion and low perfusion, alongside its performance in outcome studies, SET® stands out as the choice of pulse oximetry technology for clinicians and policy makers hoping to implement newborn-related screening processes—and has indeed been used in the establishment of screening guidelines used around the world.15
@Masimo | #Masimo
About Masimo
Masimo (NASDAQ: MASI) is a global medical technology company that develops and produces a wide array of industry-leading monitoring technologies, including innovative measurements, sensors, patient monitors, and automation and connectivity solutions. Our mission is to improve patient outcomes and reduce the cost of care. Masimo SET® Measure-through Motion and Low Perfusion™ pulse oximetry, introduced in 1995, has been shown in over 100 independent and objective studies to outperform other pulse oximetry technologies.2 Masimo SET® has also been shown to help clinicians reduce severe retinopathy of prematurity in neonates,16 improve CCHD screening in newborns,3 and, when used for continuous monitoring with Masimo Patient SafetyNet™ in post-surgical wards, reduce rapid response team activations, ICU transfers, and costs.17-20 Masimo SET® is estimated to be used on more than 200 million patients in leading hospitals and other healthcare settings around the world,21 and is the primary pulse oximetry at 9 of the top 10 hospitals according to the 2020-21 U.S. News and World Report Best Hospitals Honor Roll.22 Masimo continues to refine SET® and in 2018, announced that SpO2 accuracy on RD SET® sensors during conditions of motion has been significantly improved, providing clinicians with even greater confidence that the SpO2 values they rely on accurately reflect a patient’s physiological status. In 2005, Masimo introduced rainbow® Pulse CO-Oximetry technology, allowing noninvasive and continuous monitoring of blood constituents that previously could only be measured invasively, including total hemoglobin (SpHb® ), oxygen content (SpOC™), carboxyhemoglobin (SpCO® ), methemoglobin (SpMet® ), Pleth Variability Index (PVi® ), RPVi™ (rainbow® PVi), and Oxygen Reserve Index (ORi™). In 2013, Masimo introduced the Root® Patient Monitoring and Connectivity Platform, built from the ground up to be as flexible and expandable as possible to facilitate the addition of other Masimo and third-party monitoring technologies; key Masimo additions include Next Generation SedLine® Brain Function Monitoring, O3® Regional Oximetry, and ISA™ Capnography with NomoLine® sampling lines. Masimo’s family of continuous and spot-check monitoring Pulse CO-Oximeters® includes devices designed for use in a variety of clinical and non-clinical scenarios, including tetherless, wearable technology, such as Radius-7® and Radius PPG™, portable devices like Rad-67™, fingertip pulse oximeters like MightySat® Rx, and devices available for use both in the hospital and at home, such as Rad-97® . Masimo hospital automation and connectivity solutions are centered around the Masimo Hospital Automation™ platform, and include Iris® Gateway, iSirona™, Patient SafetyNet, Replica™, Halo ION™, UniView™, UniView :60™, and Masimo SafetyNet™. Additional information about Masimo and its products may be found at www.masimo.com . Published clinical studies on Masimo products can be found at www.masimo.com/evidence/featured-studies/feature/ .
ORi and RPVi have not received FDA 510(k) clearance and are not available for sale in the United States. The use of the trademark Patient SafetyNet is under license from University HealthSystem Consortium.
References
- Gunaratne CR, Hewage I, Fonseka A, Thennakoon S. Comparison of pulse oximetry screening versus routine clinical examination in detecting critical congenital heart disease in newborns. Sri Lanka J Child Health , 2021; 50(1): 04-11. DOI: http://dx.doi.org/10.4038/sljch.v50i1.9393 .
- Published clinical studies on pulse oximetry and the benefits of Masimo SET® can be found on our website at http://www.masimo.com . Comparative studies include independent and objective studies which are comprised of abstracts presented at scientific meetings and peer-reviewed journal articles.
- de-Wahl Granelli A et al. Impact of pulse oximetry screening on the detection of duct dependent congenital heart disease: a Swedish prospective screening study in 39,821 newborns. BMJ. 2009;Jan 8;338.
- Slitine N, et al. Pulse Oximetry and Congenital Heart Disease Screening: Results of the First Pilot Study in Morocco. Int J Neonatal Screen 6(53). 30 June 2020.
- Zhao et al. Pulse oximetry with clinical assessment to screen for congenital heart disease in neonates in China: a prospective study. Lancet . 2014 Aug 30;384(9945):747-54.
- Ewer AK et al. Pulse Oximetry Screening for Congenital Heart Defects in Newborn Infants (Pulseox): A Test Accuracy Study. Lancet . 2011 Aug 27;378(9793):785-94.
- de-Wahl Granelli A et al. Noninvasive Peripheral Perfusion Index as a Possible Tool for Screening for Critical Left Heart Obstruction. Acta Paediatr 2007; 96(10): 1455-9.
- Meberg A et al. First Day of Life Pulse Oximetry Screening to Detect Congenital Heart Defects. J Pediatr 2008; 152:761-5.
- Schena F et al. Perfusion Index and Pulse Oximetry Screening for Congenital Heart Defects. J Pediatr . 2017 Apr;183:74-79.
- Hamilçıkan S, Can E. Critical Congenital Heart Disease Screening With a Pulse Oximetry in Neonates. J Perinat Med. 2018 Feb 23;46(2):203-207.
- Jawin V et al. Beyond Critical Congenital Heart Disease: Newborn Screening Using Pulse Oximetry for Neonatal Sepsis and Respiratory Diseases in a Middle-Income Country. PLoS One . 2015; 10(9): e0137580.
- Tekleab AM, Sewnet YC. Role of pulse oximetry in detecting critical congenital heart disease among newborns delivered at a high altitude setting in Ethiopia. Pediatric Health Med Ther . 2019;10:83-88. https://doi.org/10.2147/PHMT.S217987 .
- Narayen IC et al. Accuracy of Pulse Oximetry Screening for Critical Congenital Heart Defects After Home Birth and Early Postnatal Discharge. J Pediatr . 2018;197:29-35.
- Oakley JL et al. Effectiveness of Pulse-Oximetry in Addition to Routine Neonatal Examination in Detection of Congenital Heart Disease in Asymptomatic Newborns. J Matern Fetal Neonatal Med . 2015;28(14):1736-9.
- Kemper et al. Strategies for implementing screening for critical congenital heart disease. Pediatrics. 2011 Nov;128(5):e1259-67. doi: 10.1542/peds.2011-1317.
- Castillo A et al. Prevention of Retinopathy of Prematurity in Preterm Infants through Changes in Clinical Practice and SpO2 Technology. Acta Paediatr. 2011 Feb;100(2):188-92.
- Taenzer A et al. Impact of pulse oximetry surveillance on rescue events and intensive care unit transfers: a before-and-after concurrence study. Anesthesiology . 2010:112(2):282-287.
- Taenzer A et al. Postoperative Monitoring – The Dartmouth Experience. Anesthesia Patient Safety Foundation Newsletter . Spring-Summer 2012.
- McGrath S et al. Surveillance Monitoring Management for General Care Units: Strategy, Design, and Implementation. The Joint Commission Journal on Quality and Patient Safety . 2016 Jul;42(7):293-302.
- McGrath S et al. Inpatient Respiratory Arrest Associated With Sedative and Analgesic Medications: Impact of Continuous Monitoring on Patient Mortality and Severe Morbidity. J Patient Saf. 2020 14 Mar. DOI: 10.1097/PTS.0000000000000696.
- Estimate: Masimo data on file.
- http://health.usnews.com/health-care/best-hospitals/articles/best-hospitals-honor-roll-and-overview .
Forward-Looking Statements
This press release includes forward-looking statements as defined in Section 27A of the Securities Act of 1933 and Section 21E of the Securities Exchange Act of 1934, in connection with the Private Securities Litigation Reform Act of 1995. These forward-looking statements include, among others, statements regarding the potential effectiveness of SET® . These forward-looking statements are based on current expectations about future events affecting us and are subject to risks and uncertainties, all of which are difficult to predict and many of which are beyond our control and could cause our actual results to differ materially and adversely from those expressed in our forward-looking statements as a result of various risk factors, including, but not limited to: risks related to our assumptions regarding the repeatability of clinical results; risks related to our belief that Masimo's unique noninvasive measurement technologies, including SET® , contribute to positive clinical outcomes and patient safety; risks related to our belief that Masimo noninvasive medical breakthroughs provide cost-effective solutions and unique advantages; risks related to COVID-19; as well as other factors discussed in the "Risk Factors" section of our most recent reports filed with the Securities and Exchange Commission ("SEC"), which may be obtained for free at the SEC's website at www.sec.gov . Although we believe that the expectations reflected in our forward-looking statements are reasonable, we do not know whether our expectations will prove correct. All forward-looking statements included in this press release are expressly qualified in their entirety by the foregoing cautionary statements. You are cautioned not to place undue reliance on these forward-looking statements, which speak only as of today's date. We do not undertake any obligation to update, amend or clarify these statements or the "Risk Factors" contained in our most recent reports filed with the SEC, whether as a result of new information, future events or otherwise, except as may be required under the applicable securities laws.
View source version on businesswire.com: https://www.businesswire.com/news/home/20210322005184/en/
Link:
About Business Wire
Subscribe to releases from Business Wire
Subscribe to all the latest releases from Business Wire by registering your e-mail address below. You can unsubscribe at any time.
Latest releases from Business Wire
Celltrion receives positive CHMP Opinion for SteQeyma™ (ustekinumab biosimilar) autoinjector18.12.2025 03:41:00 CET | Press release
SteQeyma™45mg and 90mg solution for injection via autoinjector (pre-filled pen) receives positive CHMP opinion, which will facilitate subcutaneous administration in patients with plaque psoriasis, psoriatic arthritis (PsA) and Crohn’s disease (CD)1The new autoinjector option increases convenience, enhances individual patient experience and expands administration options Celltrion, Inc. today announced that the Committee for Medicinal Products for Human Use (CHMP) of the European Medicines Agency (EMA) has adopted a positive opinion of autoinjector of SteQeyma™, a biosimilar to Stelara® (ustekinumab), for the treatment of plaque psoriasis, psoriatic arthritis (PsA) and Crohn’s disease (CD). The positive CHMP opinion is for SteQeyma autoinjector in 45mg/0.5mL and 90mg/1mL, expanding the currently approved SteQeyma™ presentation, which includes 45mg/0.5mL, 90mg/1mL in a pre-filled syringe and 45mg/0.5mL in a vial for subcutaneous injection, as well as 130mg/26mL concentrate for solution f
Megaport Expands into India, Accelerating Global Growth with Extreme IX Acquisition18.12.2025 02:15:00 CET | Press release
Through the Extreme Exchange (IX) acquisition, Megaport gains seven Internet Exchanges and access to 40+ data centres across India’s fastest-growing digital hubs. Megaport Limited (ASX: MP1) (“Megaport”), the world’s leading Network as a Service (NaaS) provider, today announced the acquisition of Extreme IX,India’s leading Internet Exchange operator, from Extreme Labs, a Bulgaria-headquartered software and network engineering company that incubated the Extreme IX platform. The acquisition expands Megaport’s global platform into one of the world’s fastest-growing digital infrastructure markets and supports the Company’s strategy to deliver scalable, high-performance connectivity services across APAC. The acquisition establishes Megaport’s presence across seven Internet Exchanges in major Indian metros: Delhi, Kolkata, Hyderabad, Chennai, Bengaluru, Mumbai, and Pune, connecting 40+ data centres and more than 400 customers. It also accelerates Megaport’s planned market entry by nearly thr
IonQ and QuantumBasel Expand Long-Term Partnership in Next-Generation Quantum Systems17.12.2025 22:10:00 CET | Press release
Extension solidifies QuantumBasel as IonQ’s Innovation Center in Europe; adds IonQ Tempo and next-generation system to advance quantum commercialization IonQ (NYSE: IONQ), the world’s leading quantum company, today announced an expanded agreement with QuantumBasel, the quantum initiative of uptownBasel, Switzerland’s innovation campus. The extended contract grants QuantumBasel ownership of its existing IonQ Forte Enterprise system and secures ownership of a next-generation Tempo system. This new agreement brings the total deal value of the QuantumBasel and IonQ partnership to over $60 million and extends IonQ’s on-site presence in Switzerland four more years, continuing through 2029. QuantumBasel is IonQ’s official Innovation Center in Europe, serving as a hub for European industry, academia, and research institutions to explore practical quantum computing applications and access IonQ’s latest enterprise-grade systems. “Our extended partnership with QuantumBasel represents a cornerston
Suzano Starts Up New Production Line, Boosting Its Fluff Pulp Capacity by 400%17.12.2025 21:50:00 CET | Press release
A R$490 million investment expands the supply of raw material used in the production of absorbent items Suzano, the world’s largest pulp producer, has commenced operations this week at its new fluff pulp production line located in its Limeira unit in Brazil’s São Paulo state. This R$490 million investment increases Suzano’s total fluff pulp production capacity by more than 400%, from 100,000 to 440,000 tonnes per year. The project involved converting the existing pulp line at the Limeira unit into a flexible machine, capable of producing both Eucafluff® and market pulp. Eucafluff® is used in the production of absorbent and personal hygiene products, such as baby and adult diapers, sanitary pads and pet pads. Then market pulp is supplied for making products including toilet paper, printing and writing papers, and paper packaging. Launched in 2015, Eucafluff® is the world’s first fluff pulp made from eucalyptus, delivering unique advantages like enhanced softness and flexibility, which t
SES Acknowledges Moody’s Rating Action and Reiterates Deleveraging Commitments17.12.2025 21:36:00 CET | Press release
SES S.A. (“SES” or the “Company”), a leading space solutions company, acknowledges the credit rating action announced by Moody’s Investor Service today, which follows the release of SES’ Q3 2025 results and Intelsat integration update. SES management reiterates that the Company continues to execute on its strategy with a clear plan to strengthen its key credit metrics over time. Consistent with this plan, it remains management’s intention to de-lever and return to credit metrics that are commensurate with investment grade, with a policy objective of reducing adjusted net leverage1 to at least 3.0x or below. Today’s rating action does not change the Company’s ability to operate its business, serve customers, or execute its strategic plan. SES maintains a balanced weighted average debt maturity profile of approximately five years, and the rating action from Moody’s is not expected to have a material impact on the interest payable under the Company’s existing debt facilities. SES also ben
In our pressroom you can read all our latest releases, find our press contacts, images, documents and other relevant information about us.
Visit our pressroom